Portable Gas Detector Testing And Calibration MANUFACTURED BY STATUS SCIENTIFIC CONTROLS Testing And Calibration of Portable Gas Detectors. 

In order to ensure the safe and accurate operation of Portable Gas Detectors, they need to be periodically calibrated against reference test gases that are traceable to National Physical Laboratory standards. The recommended calibration period for Status Scientific Controls gas detectors is six months although the frequency may need to be increased where continual exposure to the target gases is experienced. 
 
Between calibrations it is also recommended that the response of the detector is checked using an appropriate test gas cylinder. These cylinders are available in 'large aerosol' form and may be supplied individually or as part of a Field Calibration Kit. 
 
Status Scientific Controls offers a range of support equipment, including compressed gases, for the customer to carry out both calibration and response testing of the Portable Gas Detectors. 
Autocalibration Unit (Stock No.SS081) 
 
The Autocalibration unit is suitable for customers who own and calibrate a fleet of Portable Gas Detectors. It operates in conjunction with a PC running purpose designed calibration software provided by Status Scientific Controls. The use of high pressure compressed gas cylinders is required with this unit - these are available from a number of specialist companies who provide a certificate of composition traceable to National Physical Laboratory Standards with each cylinder. 
Warning - This unit is for use only within a safe area where potentially explosive gases do not exist. 
PC Interface Or PC Interface/Charger Units (Stock No.SS103) 
 
A PC Interface Unit or PC Charger/Interface Unit provides cost effective method of calibrating the Portable Gas Detector. It operates in conjunction with a PC running purpose designed calibration software provided by Status Scientific Controls. Test gas cylinders, regulators and sampling adaptors are required in order to present known reference gas levels to the instrument. These are available from Status Scientific Controls either individually or in kit form. 
Warning - This unit is for use only within a safe area where potentially explosive gases do not exist. 
Test Gases For Calibration And Response Testing 
 
Various test gas mixtures can be supplied together with cylinder regulators, tubing and gas-inlet sampling adaptors. 
see the order codes below for details. 
Stock No. SS081 - Autocalibration Unit c/w software. 
Stock No. SS102 - PC Interface Unit. 
Stock No. SS103 - PC Interface/Charger Unit. 
Stock No. SS105 - Field Calibration Kit. 
Stock No. SS108 - Metal Sample Flow Adaptor & Tubing. 
Stock No. SS111 - PTFE Sample Flow Adaptor & Tubing (for absorbent gases e.g. Chlorine, Sulphur Dioxide). 
Stock No. SS219 - Regulator kit c/w sample adaptors & tubing. 
Stock No. 
Description 
SS220 
Methane 25% LEL, balance air 
SS221 
Methane 50% LEL, balance air 
SS222 
Propane 20% LEL, balance ai 
SS226 
Carbon Monoxide 50ppm, balance air 
SS227 
Carbon Monoxide 200ppm, balance air 
SS229 
Hydrogen Sulphide 25ppm, balance Nitrogen 
SS230 
Sulphur Dioxide 10ppm, balance Nitrogen 
SS231 
Compressed Air 20.9% 
SS232 
Propane 50% LEL, balance air 
SS233 
Carbon Dioxide 2%, balance air 
SS234 
Methane 99.9% volume 
SS235 
Methane 30% LEL, balance air 
SS236 
Carbon Monoxide 65ppm, Methane 1.5% volume, Hydrogen Sulphide 25ppm, Oxygen 18.5%, balance Nitrogen 
SS237 
Hydrogen Sulphide 80ppm, balance Nitrogen 
SS238 
Hydrogen Sulphide 30ppm, Methane 2.5% volume, Oxygen 18.5%, balance Nitrogen 
SS239 
Chlorine 10ppm, balance Nitrogen 
SS240 
Carbon Monoxide 65ppm, Methane 1.5% volume, Oxygen 18.5%, balance Nitrogen 
SS241 
Hydrogen 1% volume, balance air 
SS242 
Nitrogen Dioxide 10ppm, balance air 
SS243 
Hydrogen Chloride 10ppm, balance Nitrogen 
SS244 
Hydrogen Cyanide 10ppm, balance Nitrogen 